An investigation into the viability of an infrared diagnostic instrument for measurement of CO2 isotope ratios in breath

An investigation into the viability of an infrared diagnostic instrument for measurement of CO2 isotope ratios in breath
An investigation into the viability of an infrared diagnostic instrument for measurement of CO2 isotope ratios in breath
Stable CO2 isotope ratio breath tests are established as a valuable tool in diagnostic and investigative medicine, with the potential to become more prominent in the future. The instrument conventionally used to measure the very small changes involved is an Isotope Ratio Mass Spectrometer however, the expense and complexity of such an instrument severely restricts the widespread and routine use of isotopic breath tests. Hence, to realise their full potential an alternative technique is required which is reliable, insensitive to environmental and component fluctuations, uncomplicated and affordable. We present a system that satisfies these criteria using broadband, non-dispersive, ground-state absorption infrared spectroscopy. Two isotopically distinct channels are created and their basal path length ratio recorded at the condition of transmitted intensity equilibrium. The change in channel path length ratios required to restore equilibrium in the 13C-enriched breath sample is directly related to the change in 13CO2 / 12CO2 concentration ratio. The system's novelty lies in this negative feedback loop balancing the signal by means of adjusting the path length of one of the channels.

There is little evidence in the literature on infrared breath measurements that the possibility of interference from coincident infrared active breath trace compounds or various spectral effects that could lead to spurious results have been adequately assessed. Therefore, prior to the construction of a prototype instrument based upon this technique the question of its validity must be addressed. Furthermore, in addition to any instrument presenting a viable alternative it should also offer better reliability, long term stability and ideally be of lower cost than other emergent infrared instruments. Significant emphasis has therefore been placed upon evaluation of possible interferents, with the intention of providing an unambiguous assurance of the measurement's validity over a range of conceivable operating conditions and establishing operating tolerances that improve reliability over other infrared techniques. Also the prospect of using this technique to perform alternative isotope ratio breath tests was explored with the feasibility of using a novel filtering technique to enable measurement of 18O12C16O / 13C16O2 being investigated further.

These aims were accomplished through the thorough evaluation of literature on breath trace compound's infrared spectra and by a series of theoretical computer instrument simulations, using detailed modelling of a breath sample's CO2 spectroscopy, capable of identifying, evaluating and quantifying the risks posed to a reliable measurement of 13CO2 / 12CO2 due to various spectral effects. The breath trace compound analysis revealed that 13C16O2 / 12C16O2 ratios can confidently be measured for isotopic breath tests using an instrument based on infrared absorption, the position of CO2's ν3 absorption band precluding any discernible risk. From the instrument simulations it has been possible to establish operating tolerances required to avoid or limit the generation of a spurious results and determine parameters necessary for the instrument's construction. Thus, we conclude that through the merits of its design and compliance to operating tolerances established the instrument described presents a viable alternative for providing highly accurate, reliable and low cost breath tests capable of being performed outside of a laboratory environment.
University of Southampton
Mansfield, Colin David
2354cf62-85aa-497f-8a77-b0705e608bf3
Mansfield, Colin David
2354cf62-85aa-497f-8a77-b0705e608bf3
Rutt, Harvey
e09fa327-0c01-467a-9898-4e7f0cd715fc

Mansfield, Colin David (2000) An investigation into the viability of an infrared diagnostic instrument for measurement of CO2 isotope ratios in breath. University of Southampton, Department of Electronics and Computer Science, Doctoral Thesis, 274pp.

Record type: Thesis (Doctoral)

Abstract

Stable CO2 isotope ratio breath tests are established as a valuable tool in diagnostic and investigative medicine, with the potential to become more prominent in the future. The instrument conventionally used to measure the very small changes involved is an Isotope Ratio Mass Spectrometer however, the expense and complexity of such an instrument severely restricts the widespread and routine use of isotopic breath tests. Hence, to realise their full potential an alternative technique is required which is reliable, insensitive to environmental and component fluctuations, uncomplicated and affordable. We present a system that satisfies these criteria using broadband, non-dispersive, ground-state absorption infrared spectroscopy. Two isotopically distinct channels are created and their basal path length ratio recorded at the condition of transmitted intensity equilibrium. The change in channel path length ratios required to restore equilibrium in the 13C-enriched breath sample is directly related to the change in 13CO2 / 12CO2 concentration ratio. The system's novelty lies in this negative feedback loop balancing the signal by means of adjusting the path length of one of the channels.

There is little evidence in the literature on infrared breath measurements that the possibility of interference from coincident infrared active breath trace compounds or various spectral effects that could lead to spurious results have been adequately assessed. Therefore, prior to the construction of a prototype instrument based upon this technique the question of its validity must be addressed. Furthermore, in addition to any instrument presenting a viable alternative it should also offer better reliability, long term stability and ideally be of lower cost than other emergent infrared instruments. Significant emphasis has therefore been placed upon evaluation of possible interferents, with the intention of providing an unambiguous assurance of the measurement's validity over a range of conceivable operating conditions and establishing operating tolerances that improve reliability over other infrared techniques. Also the prospect of using this technique to perform alternative isotope ratio breath tests was explored with the feasibility of using a novel filtering technique to enable measurement of 18O12C16O / 13C16O2 being investigated further.

These aims were accomplished through the thorough evaluation of literature on breath trace compound's infrared spectra and by a series of theoretical computer instrument simulations, using detailed modelling of a breath sample's CO2 spectroscopy, capable of identifying, evaluating and quantifying the risks posed to a reliable measurement of 13CO2 / 12CO2 due to various spectral effects. The breath trace compound analysis revealed that 13C16O2 / 12C16O2 ratios can confidently be measured for isotopic breath tests using an instrument based on infrared absorption, the position of CO2's ν3 absorption band precluding any discernible risk. From the instrument simulations it has been possible to establish operating tolerances required to avoid or limit the generation of a spurious results and determine parameters necessary for the instrument's construction. Thus, we conclude that through the merits of its design and compliance to operating tolerances established the instrument described presents a viable alternative for providing highly accurate, reliable and low cost breath tests capable of being performed outside of a laboratory environment.

Text
Mansfield 2000 thesis 1480T - Version of Record
Available under License University of Southampton Thesis Licence.
Download (14MB)

More information

Published date: June 2000
Organisations: University of Southampton, Optoelectronics Research Centre, Electronics & Computer Science

Identifiers

Local EPrints ID: 15497
URI: http://eprints.soton.ac.uk/id/eprint/15497
PURE UUID: 1c48d018-2ade-4620-9043-9ee7beecad2c

Catalogue record

Date deposited: 09 Jun 2005
Last modified: 15 Mar 2024 05:40

Export record

Contributors

Author: Colin David Mansfield
Thesis advisor: Harvey Rutt

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×